Summary
Type: SPE
Application #: 303870
Catalog#: 730014 - SPE Cartridges, Chromabond, C18 ec, 6ml, PP, 500mg, 30/pk
Detection: HPLC according Application No.: 110050 and 110730
Conditions: Column conditioning:
5 ml acetontrile, and then 10 ml dest. water
Sample application:
Apply the prepared sample on the column
Washing:
10 ml dest. water
Elution:
2 ml 0.1% triethylamine in acetonitrile (v/v)
Authors: Macherey-Nagel
Source: SPE department